EAS Timeout

A setting of 0 (000) disables this feature.
NOTE
Product Reference Guide
Specifies the amount of time that an EAS (Electronic Article Survel-
lance) signal is held in its active state for a good read indication.
To set the EAS Timeout:
1. Scan the SWITCH bar code.
2. Scan the bar code, SET EAS TIMEOUT on page 147 below.
You'll need to cover any unused bar codes on this and the fac-
ing page to ensure that the scanner reads only the bar code
you intend to scan.
3. Scan the appropriate digits from the keypad in Appendix C
that represent the desired EAS Timeout duration. The select-
able range is 0-255, which is the timeout in 10-millisecond
increments. Pad all single and double digit numbers with lead-
ing zeroes to yield a three-digit entry (000-255).
Examples:
000 = EAS timeout disabled
001 = 10ms
005 = 50ms
040 = 400ms
250 = 2,500ms (2.5 seconds)
4. Scan the SWITCH bar code to exit Programming Mode.
SET EAS TIMEOUT
EAS Features
147