Error Indicators; Error Code Table - Beckman Coulter StatSpin Express 4 M510 Instructions For Use Manual

Centrifuge
Table of Contents

Advertisement

System Overview

Error Indicators

Table 2.1
Note
Buttons should be depressed with finger tips only. Never press buttons with a sharp
object such as a pen, screwdriver, centrifuge insert, fingernail, etc. The buttons are
membrane switches designed to be activated by finger actuation. Use with any hard or
sharp object can cause damage to the tactile layer of the button, rendering the button
unstable and prone to premature failure.
Error Indicators
A flashing red Error or Service indicator on the control panel signals a StatSpin Express 4
error condition.
Use the combination of green cycle time indicators that are flashing to identify the cause of
the error condition. For more information, refer to
To clear the error condition, correct the problem and press the Stop or Open button.
1. Error or Service light flashing

Error Code Table

Table 2.2
Cycle Time Indicator Combination
2-2
Express 4 Operator Controls (Continued)
Symbol
Error Code Table
Unlock
Unlocked position for shield.
Error
Problem
Failure to reach speed in 30
seconds.
Description
Indicators.
Solution
Confirm no obstruction in
bowl area, contact customer
service.
55-006276-001CC