GE H
DRAFT
EALTHCARE
D
5460683, R
IRECTION
EVISION
2-2-4
Probe Environmental Requirements
Table 2-5
Temperature
Humidity
Pressure
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
NOTICE
0 TO + 55 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
!! NOTICE:
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
2-2-5
Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Have two people available to deliver and unpack the LOGIQ™ S7 ultrasound system.
CAUTION
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
!! CAUTION:
result in personal injury and/or damage to the system.
3
DRAFT (J
2D Probe Environmental Requirements
Operational
o
10
- 35
for 2D probe
o
o
18
- 30
for 4D probe
10 - 80% (non-condensing) for 2D probe
30 - 80% (non-condensing) for 4D probe
700 - 1060hPa
Chapter 2 - Site Preparation
25, 2014)
UNE
o
C
F
LOGIQ™ S7 E
/P
XPERT
RO
S
M
ERVICE
ANUAL
2-5
Need help?
Do you have a question about the LOGIQ S7 Expert and is the answer not in the manual?