5460683, R
IRECTION
EVISION
10.)A message appears, and then presses CTRL-ALT-DEL to reboot the system.
11.)At the first system reboot after loading the software, the OS will reboot the system automatically.
12.)After the reboot, the installation of system (Windows XP Embedded) will be fully completed.
8-3-7-2

System Software - Installation Procedure (R2.x.x or later)

1.) Prepare the Base System Software DVD.
2.) Power ON the scanner and press "DEL" to enter BIOS
3.) Type "Polaris" for password.
4.) Go to "Boot" in BIOS menu and change the setting value from "SATA 2" to "SATA 0" in Boot Option
#1
5.) Go to "Save & Exit" in BIOS menu and select "Save Changes and Exit". Then, the system will re-
3