Type Bf Patient Leakage Current - Mains To Probe - GE LOGIQ S7 Expert Service Manual

Expert ultrasound system
Hide thumbs Also See for LOGIQ S7 Expert:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
5460683, R
IRECTION
EVISION
10-8-8

Type BF Patient Leakage current - Mains to Probe

Reference the procedure in the IEC 60601-1. Measure leakage current flow from mains to Probe.
Electric Shock Hazard.
DANGER
Line voltage is applied to Probe during this test. To avoid possible electric shock hazard, the
!! DANGER:
system being tested must not be touched by patients, users or anyone during testing.
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
CAUTION
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
!! CAUTION:
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.
Table 10-19 Typical Data Sheet for Type BF Patient Leakage Current - Mains to Probe
Tester Neutral
Closed
Closed
NOTE:
It is not necessary to test each lead individually or power condition combinations as required in
previous tests.
10-30
3
DRAFT (J
Tester Polarity Switch
Normal
Reversed
Section 10-8 - Electrical Safety Tests
25, 2014)
UNE
Ground Switch
Closed
Closed
LOGIQ™ S7 E
/P
XPERT
RO
S
M
ERVICE
ANUAL
Measured leakage current

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LOGIQ S7 Expert and is the answer not in the manual?

This manual is also suitable for:

Logiq s7 pro

Table of Contents

Save PDF