Printer Malfunction

configured properly?
impossible by manual
If unsuccessful replace printer
Return to normal scanning
3
DRAFT (J
Printer malfunction
yes
Is printer
properly
connected?
yes
Is printer
loaded with paper?
yes
Is printing media
(ink, cartridge, etc.)
empty?
yes
Are the P-keys
yes
Is printing also
operation?
yes
Perform Printer Test
Print Test page
Figure 7-15 Printer Malfunction
Chapter 7 - Diagnostics/Troubleshooting
25, 2014)
UNE
Check all signal and power supply cable
no
connections between Printer and Scanner
Load paper by using the correct format.
no
Check paper size (small, large, etc.)
Check paper tray to fit with the used paper format
Depending on the Printer Type
no
replace ink, cartridge etc. color
Check programmed P-keys and
no
re-configure them if not correct
no
Check remote cable