Location, Region and Language
1.) From the touch panel, select Utility -> System-> General.
2.) Record all settings in
Table 8-17
Parameter
Hospital
Department
Preset Region
Language
Units
8-3-6

System Software - Installation Procedure

Two separate methods of software installation is offered in LOGIQ S7 Expert/Pro.
A.) Software Load from DVD
B.) Software Re-Load from Repository
Situation
8-3-7
Software Installation from DVD
8-3-7-1
System Software - Installation Procedure (R1.x.x)
1.) Prepare the Base System Software DVD.
2.) Power ON the scanner.
3.) Press DVD-Drive eject button.
4.) Place the Base System Software DVD on the tray.
5.) Wait few moments for DVD drive to read DVD contents.
3
DRAFT (J
Figure 8-13 Location, Region and Language
Table
8-17.
Record settings from Location
• Harddrive is new
• Reload from Repository fails
• Wish to perform "Clean Install"
Perform Software Load from DVD
Refer to Section 8-2-6
Chapter 8 - Replacement Procedures
25, 2014)
UNE
Value
• Re-installation for application or
operating system issues
Perform Software Reload
Refer to Section 8-2-7
LOGIQ™ S7 E
/P
XPERT
RO
S
M
ERVICE
ANUAL
8-19