New Unit - GE LOGIQ S7 Expert Service Manual

Expert ultrasound system
Hide thumbs Also See for LOGIQ S7 Expert:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
5460683, R
IRECTION
EVISION
10-9-5

New Unit

If the leakage current measurement tests fail on a new unit and if situation cannot be corrected, submit
a Safety Failure Report to document the system problem. Remove unit from operation.
ELECTRICAL SAFETY
Outlet and Ground Continuity test
Table 10-20
Outlet/Ground Continuity
Outlet (correct ground & wiring config.)
System Ground Continuity
Earth leakage current test
Table 10-21
Tester Neutral
Closed
Closed
Open
Open
Enclosure leakage current test
Table 10-22
Tester
Neutral
Closed
Closed
Closed
Closed
Open
Open
10-32
3
DRAFT (J
Tester Polarity
Switch
Normal
0.3mA(USA)/0.5mA(Others)
Reversed
0.3mA(USA)/0.5mA(Others)
Normal
1mA
Reversed
1mA
Tester
Protective
Max Value
Polarity
Earth
Switch
Switch
0.1mA(USA)/
Normal
Closed
0.1mA(Others)
0.3mA(USA)/
Normal
Open
0.5mA(Others)
0.1mA(USA)/
Reversed
Closed
0.1mA(Others)
0.3mA(USA)/
Reversed
Open
0.5mA(Others)
0.5mA(USA)/
Normal
Closed
0.5mA(Others)
0.5mA(USA)/
Reversed
Closed
0.5mA(Others)
Section 10-9 - When There's Too Much Leakage Current...
25, 2014)
UNE
Max Value
Value
Allowed
Measured
N/A
N/A
0.2 Ohm
Max Value Allowed
Enter Name of tested peripheral here:
Test 1
Test 2
Allowed
Wheel
LCD
Enter Name of tested peripheral here:
LOGIQ™ S7 E
S
ERVICE
OK?
Value
OK?
Measured
Optional
Optional
OK?
Test 3
Test 4
/P
XPERT
RO
M
ANUAL
Comments
Comments
Comments