Modifying Contents Of Eeprom (Comexpress Vpd) - GE LOGIQ S7 Expert Service Manual

Expert ultrasound system
Hide thumbs Also See for LOGIQ S7 Expert:
Table of Contents

Advertisement

GE H
DRAFT
EALTHCARE
D
5460683, R
IRECTION
EVISION
8-21-6-3

Modifying contents of EEPROM (ComExpress VPD)

System reads off Console Serial Number from EEPROM and match that against installed option string.
Therefore, without having correct serial number in EEPROM (ComExpress VPD).
1.) Select "COMEXPRESS" from Device Grid (right pane).
2.) Double Click value for "System Serial Number".
3.) Type in console's serial number.
4.) Click "Write Dirty Fields" button on right pane.
NOTE:
This procedure can be applied to other components with VPD, for example, for modifying GFS FuncRev
value. However, it is unlikely need for such event arises, as these VPD components are hardware-
integrated into boards.
8-21-7
Functional Check
Board Diagnostics
3
DRAFT (J
Figure 8-114
Test
Chapter 8 - Replacement Procedures
25, 2014)
UNE
COMEXPRESS
Refer to
Section Section 4-4
LOGIQ™ S7 E
/P
XPERT
RO
S
M
ERVICE
ANUAL
Note
8-101