  • Page 1 ÄKTAcrossflow ™ Site Preparation Guide...
  • Page 2 Table of Contents Table of Contents Introduction ......................Site preparation overview .................. Site preparation checklist .................. ÄKTAcrossflow Site Preparation Guide 29-1396-59 AA...
  • Page 3 Chapter 3. Return the completed Site preparation checklist to the GE service office in your region. You will find E-mail address and fax number to the GE service office in your region under: www.gelifesciences.com/contact . If your region is not listed, select the nearest region.
  • Page 4 The site preparation tasks listed in the following table can be divided between two or three people from the customer. It is recommended that the same people assist the GE service personnel during the in- stallation process later on. Personnel...
  • Page 5 2 Site preparation overview Site preparation overview Introduction Before the product is delivered, you must prepare the site according to the instructions in this chapter. Select the site Select a site that complies with national and international requirements for biosafety. WARNING Spread of biological agents.
  • Page 6 2 Site preparation overview Note: The computer and screen have their own specifications and requirements. Make sure to follow these when connecting them. WARNING Access to power cord. Do not block access to the power cord. The power cord must always be easy to disconnect. Dimensions and weight The following illustration shows the dimensions of the ÄKTAcrossflow system.
  • Page 7 2 Site preparation overview Environmental requirements The installation site must comply with the following specifications: Parameter Specification Ambient temperature, operation 4°C to 40°C Relative humidity 20% to 95%, non-condensing Atmospheric pressure 840 to 1060 mbar Electrical requirements Parameter Specification Supply voltage 100 to 240 V AC Frequency 50/60 Hz...
  • Page 8 2 Site preparation overview The ÄKTAcrossflow system consists of a ÄKTAcrossflow system separation unit, a con- nected computer including a flat-screen monitor, and UNICORN control system. The following table specifies the computer requirements. Parameter Specification ® Operating system Windows 7 Professional, 32-bit or 64- bit or Windows XP Professional, SP3 Control system...
  • Page 9 2 Site preparation overview Item Specification Acetone 1000 ml 1 M sodium chloride with 0.4% acetone and 15 mM phosphoric acid added Sodium chloride Phosphoric acid Ethanol Sodium hydroxide 10 mM Note: All reagents should be filtered and degassed, and all reagents should have relevant certificates.
  • Page 10 Complete the checklist in this chapter to make sure that all necessary preparations have been made for the installation of ÄKTAcrossflow. You must send the completed checklist to a GE office before GE service personnel can install ÄKTAcrossflow. See E-mail addresses and fax numbers in Chapter 1.
  • Page 11 3 Site preparation checklist Requirement Comment/specify Space requirements Space is available as specified. Environmental requirements The proper environmental conditions are fulfilled as specified. Ambient temperature Relative humidity Atmospheric pressure Electrical requirements Electrical requirements are fulfilled as specified. Three (3) grounded wall outlets Computer requirements A computer has been purchased and received Has an administrator account been creat-...
  • Page 12 3 Site preparation checklist Requirement Comment/specify High quality water, distilled, 3 liters Ethanol 20% Acetone 0.4% (For 1000 ml 1 M sodium chloride with 0.4% acetone and 15 mM phosphoric acid added) Sodium chloride 1 M (For 1000 ml 1 M sodium chloride with 0.4% acetone and 15 mM phosphoric acid added) Phosphoric acid 15 mM...
  • Page 13 3 Site preparation checklist Customer contact information Date (dd-mmm-yyyy) Company/Customer name Company/Customer representative Name Address Email Phone number Customer experience level Beginner Intermediate Advanced ÄKTAcrossflow Site Preparation Guide 29-1396-59 AA...
  • Page 14 Sweden All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms www.gelifesciences.com/contact and conditions is available on request. Contact your local GE Healthcare repre- sentative for the most current information.