4.2.14 Min Pressure
The Min pressure setting is only used when Target Volume is activated.
Min pressure defines the lower pressure limit down to where the Vivo 3 can decrease
the pressure to maintain the set Target Volume. If the actual volume is above Target
Volume at Min pressure, the Vivo 3 will continue to ventilate at this Min pressure
setting.
Property
Value range
Default value
Setting resolution
Available in modes
4.2.15 Ramp Up
The ramp up setting defines a ramp time for increasing the airway pressures.
At the beginning of Ramp Up time, IPAP will start at 2 cmH
EPAP.
Property
Value range
Default value
Setting resolution
Available in modes
Doc. 007232 Y-2
Description
4 cmH
O to 20 cmH
O.
2
2
Same as the value for IPAP
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
• PCV+A
• PSV
• S
• S/T
• T
Description
• Off
• 10 min to 60 min
Off
10 min
• PCV+A
• PSV
• S
• S/T
• T
• CPAP
Treatment Functions and Settings 59
O
2
O
2
O above the initially set
2
Vivo 3 Clinician's Manual