Data Parameters

SETTING/VALUE
Cough Modes
Theraputic Modes
(Not ventilation)
Device Modes
Insufflation Pressure
Exsufflation Pressure
Insufflation time (Ti)
Exsufflation time (Te)
Pause time
Oscillation Frequency
Oscillation amplitude
67
Technical Specifications
Clearway 2 User Manual
RANGE/PERFORMANCE
• Manual Mode
• Basic Auto Mode
• Program Auto Mode
• TreatRepeat (as a saved profile)
• IPPB Mode
• NIV Mode
• Clinical (unlocked)
• Home (locked)
3 to 70 cmH
O
2
Resolution 1 cmH
Tolerance: ± 5% FS
-3 to -70 cmH
O
2
Resolution 1 cmH
Tolerance: ± 5% FS
0.5 – 5.0 seconds
Resolution 0.1 second
Tolerance ± 0.25 seconds
0.5 - 5.0 seconds
Resolution 0.1 second
Tolerance ± 0.25 seconds
0.0 - 9.0 seconds (mode dependant)
Resolution 0.1 second
Tolerance ± 0.25 seconds
0 – 20 Hz
Resolution 1 Hz
Tolerance ± 1 Hz
0 - 10 cmH
O
2
Resolution 1 cmH
Tolerance ± 1 cmH
O
2
O
2
O
2
O
2
DOC-001000 Ver 6