Risk of Fire
Do not use aerosols or solvents close to the oxygen supply, even when the oxygen
supply is shut off.
CAUTION!
Supplemental oxygen flow must not exceed 30 l/min or 100 kPa.
CAUTION!
Supplemental oxygen is added before the volume measurement sensor and thereby
included in the measurements. However, the oxygen concentration still has influence
on the volume measurement of the delivered air.
This measurement is based on a normal oxygen concentration of 21%. If the oxygen
concentration is higher, the actual inspired volume will deviate from the monitored
volume as follows:
•
40% oxygen concentration: -2.5% deviation
•
60% oxygen concentration: -5% deviation
•
80% oxygen concentration: -7.5% deviation
2.9
Mobile Use — Warning and Precautions
This section applies if using the Vivo 2 during transit, for example on a wheel chair or in
a car.
WARNING!
Risk of Suffocation
Do not use the attachable humidifier during mobile use.
Due to movements, water spillage from the humidifier or condensed water may flow
to the patient and cause suffocation.
WARNING!
Risk of Electric Shock
Do not use the attachable humidifier during mobile use.
Internal water spillage may cause electric shocks and may damage the device.
28
Safety Information
User manual
Doc. 007229 D-3