Recommended separation distances between external power
conductors and the Clearway 2
Rated Maximum Current in
Conductor (A)
1
10
30
For conductors rated at a maximum current not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation d=I/2πH, where I is the maximum current rating of the conductor
in amperes (A) according to the transmitter manufacturer; H is the Clearway
2 immunity compliance level to electromagnetic fields in the 50-60 Hz
frequency span (30 A/m).
Portable RF devices (e.g. mobile phones) should be used no
closer than 30 cm (12 inches) to any part of the ME Equipment,
including any cables

Clearway 2 in other countries

If you are traveling to a country with a line voltage which is different to where
you were issued the device, then a different power cord or an international plug
adaptor may be required to ensure your power cord is compatible with the
power outlet of the country where you are travelling.
77
Technical Specifications
Clearway 2 User Manual
Separation Distance (M)
50-60 Hz
d= I/2πH= I/188
0.005
0.05
0.16