4.8 Disabling exposure key

Enter the service mode according to the instruction given in section 4.2 "How to enter/exit
service mode" on page 19.
Press the parameter adjustment up key until the parameter number 18 appears on the kV
display.
18
The number indicating the mode of exposure key operation is shown on the time display. 0
0
= exposure key normal operation, 1 = exposure key operation disabled. The factory
default is 0.
Press the Select key until the number starts to blink, and the value can now be changed
with the parameter adjustment keys.
Select key
Accept the new setting by pressing the Select key.
To exit the service mode press the Mode key for 2 seconds.

4.9 Automatic identification of cascade card

Enter the service mode according to the instruction given in section 4.2 "How to enter/exit
service mode" on page 19.
Press the parameter adjustment up key until the parameter number 19 appears on the kV
display.
19
The value indicating the type of the cascade card is shown on the time display. Value <
108 equals an old card and value  108 equals a new card. The type of the cascade card
108
affects the adjustment functionality of the preheat and mA.
Technical manual
Parameter adjustment up -key
Parameter adjustment keys
Parameter adjustment up -key
4 SERVICE MODE
Planmeca ProX
23