Ceph Mode Beam Right Position; Too Left/Right - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

3.1.a Ceph Mode Beam Right Position
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Press Test on the bottom and take an exposure. Beam should be hitting
from the center to the right on the screen. (Figure 125)
Figure 125
If not aligned, adjust Ceph mode C-arm rotation.
3.1.a.i Too Left/Right
If the beam is too left, move the Ceph mode beam right
position to the right. (Figure 126)
Figure 126
- 70 -
70