Test 3: Power Measurement

Test 3: Power Measurement
This set of tests verifies the power-measurement specifications of
Agilent 53147A/148A/149A's Power Meter.
Equipment Required
Agilent 11683A Range Calibrator
Agilent 11730A Power Sensor Cable (provided with 53147A/148A/149A)

Power Measurement Accuracy Test

To ensure that the test results are valid, the instrument and the test
equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the
equipment to stabilize.
These tests are appropriate for the Agilent 53147A, 53148A, and 53149A.
Connect the Agilent 11730A Power Sensor Cable between Agilent 11683A
Range Calibrator's output connector and the Agilent 53147A, 53148A, or
53149A's Power Meter INPUT connector.
Agilent
53147A/48A/49A
Agilent
Figure 1-6. Power Measurement Accuracy Test Setup
Press the Reset/Local key on the Agilent 53147A, 53148A, or 53149A.
Assembly Level Service Guide
Agilent
Range Calibrator
Power Input
11683A