Source Calibration Adjustment - Agilent Technologies E8362A Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A

Source Calibration Adjustment

The source calibration is used to adjust your network analyzer for a flat source power
across its full frequency range. The differences between Port 1 and Port 2 are negligible;
therefore, a Port 2 adjustment is not required.
Equipment Used for the Source Calibration Adjustment
Equipment Type
Power meter
Power sensor, 3.5 mm
Adapter, 3.5 mm (f) to 3.5 mm (f)
Power sensor, 2.4 mm
Adapter, 2.4 mm (f) to 2.4 mm (f)
Procedure
1. Connect the equipment as shown in
network analyzer and the power meter.
Figure 3-18 Setup for the Source Calibration Adjustment
2. On the System menu, point to Service, Adjustments, and then click
Source Calibration.
3. Ensure the GPIB settings are correct.
4. Click Calibrate, and then follow the instructions as they are displayed.
Service Guide E8364-90001
Models
Model or
Used With
Part Number
All
E4418B/E4419B
E8362A
E4413A
E8362A
83059B
E8363A, E8364A
8487A
E8363A, E8364A
11900B
Figure
3-18. Connect a GPIB cable between the
Tests and Adjustments
Adjustments
Alternate Model
or Part Number
E4418A/E4419A
8485A
85052-60012
None
85056-60007
3- 43

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363aE8364a

Table of Contents