Correction Constants

The analyzer stores many correction constants in non-volatile EEPROM memory. These constants enable
the analyzer to produce accurate, leveled source signals and receive clean test signals.

Storing Correction Constants

After performing any adjustment listed on
backup file on the analyzer hard disk drive by performing these steps:
• Navigate to the EEPROM Backup Utility, located at:
C:\Program Files\Agilent\Network Analyzer\Service\eebackup.exe
the program.
Run
• Click
Backup EEPROM
• Click
when the program has finished.
Exit
Service Guide E8364-90038
page 3-38
.
General Purpose Maintenance Procedures
in this manual, store the correction constants to a
Correction Constants
8-21