PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Correction Constants
The analyzer stores many correction constants in non-volatile EEPROM memory. These
constants enable the analyzer to produce accurate, leveled source signals and receive clean
test signals.
Storing Correction Constants
After performing any adjustment listed on
constants to a backup file on the analyzer hard disk drive by performing these steps:
• Navigate to the EEPROM Backup Utility, located at:
C:\Program Files\Agilent\Network Analyzer\Service\eebackup.exe
•
the program.
Run
• Click
Backup EEPROM
• Click
when the program has finished.
Exit
Service Guide E8364-90026
page 3-40
.
General Purpose Maintenance Procedures
Correction Constants
in this manual, store the correction
8-21