Correction Constants; Storing Correction Constants - Agilent Technologies N5230A Service Manual

2-port pna-l microwave network analyzers
Table of Contents

Advertisement

General Purpose Maintenance Procedures

Correction Constants

Correction Constants
The analyzer stores many correction constants in non-volatile EEPROM memory. These
constants enable the analyzer to produce accurate, leveled source signals and receive clean
test signals.

Storing Correction Constants

After performing any adjustment listed on
constants to a backup file on the analyzer hard disk drive by performing these steps:
• Navigate to the EEPROM Backup Utility, located at:
C:\Program Files\Agilent\Network Analyzer\Service\eebackup.exe
the program.
Run
• Click
Backup EEPROM
• Click
when the program has finished.
Exit
8-20
page 3-35
.
PNA Series Microwave Network Analyzers
in this manual, store the correction
Service Guide N5230-90014
N5230A