Test Procedure For The 70 Mhz If Out; Amplitude Response - Agilent Technologies ESA Series User's And Service Manual

Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

E440xBU Option H70

Test Procedure For the 70 MHz IF Out

Amplitude Response

This test measures the ESA's amplitude response as a function of frequency. The test
calculates the amplitude response relative to 50 MHz and the peak-to-peak response.
A signal is applied to the analyzer at several frequencies. At each frequency the signal
amplitude is adjusted to place the signal at approximately the same displayed amplitude.
A power meter measures the signal amplitude.
Follow the steps below to prepare for the amplitude verification of the 70 MHz IF Output
on the rear of the ESA.
1. [Preset] the E440xBU ESA and all the test equipment.
2. Run Align All Now on the ESA. Press: [System] > Alignments > Align All Now.
3. Initialize the test equipment parameters as follows.
Table 5
E440xBU ESA Signal Analyzer Setup
Parameter
Center Frequency
Span
Resolution Bandwidth
Attenuation
Frequency Reference
1. Select 0 dB since this setting will determine the
IF output level relative to the input signal.
Table 6
Synthesized Sweeper Setup
Parameter
Frequency
Function
Level
4. Zero and calibrate the power sensor.
5. Connect the power sensor/power meter to the Type-N cable that is connected to the
Source. Use an N (f) to N (f) adapter between the power sensor and the cable.
6. Adjust the Source [Amplitude] for a power meter reading of –10 dBm.
Installation Guide
Setting
1 GHz
0 Hz
1 MHz
1
0 dB
Internal
Settings
1 GHz
CW or Sine
–10 dBm
Performance Test
11