Calibration; Bench Calibration; D-To-A Adjust; Calibration Flowchart (Ebth - ABB STT04 Instruction Manual

Smart transmitter terminal
Table of Contents

Advertisement

Key
STORE THIS
CONFIGURATION
E N TE R
→ YES
ID TAGNAME
E N TE R

CALIBRATION

B EN C H
C A LIBR ATIO N
C
A
L
I
B
R
A
T
E
D -T O -A
Display
NO
READY
This section details the EBTH temperature transmitter calibration
functions of the STT04 terminal. There are two types of calibration
functions:

Bench calibration.

D-to-A adjust.

NOTES:
1.
When calibrating, a field device must be connected to the
STT04 terminal.
2.
To change calibration or configuration parameters of a field
device that is connected to an IMFBS01 module, the transmitter
must be taken off-line. This is done at the INFI 90 OPEN console.
Refer to Figure
B-2
A D JU S T LO W
A D JU S T H IG H
C A LIBR ATIO N
C A LIBR ATIO N
A P P LY 1 00 m V
S O U R C E
A R R O W
A D JU S T
K EY S
O U T PU T
TO 4 m A
C A L
M E TH O D
4 m A
C A LIBR ATIO N
E N TE R M ETE R
M E TE R VA L U E
VA LU E 4.00 0
E N TRY
Figure B-2. Calibration Flowchart (EBTH)
EBTH TEMPERATURE TRANSMITTER
Comments
Select YES.
Configuration ID tag name just configured will be in
the upper left corner if the configuration is saved. It
becomes the working configuration.
for an overview of the calibration functions.
A D JU S T LO W
A D JU S T H IG H
C A L IBR ATIO N
C A L IBR AT IO N
R E SIS TA N C E
R E SIS TA N C E
A PP LY 0 Ω
A PP LY 25 0 Ω
A D JU S T
O U TP U T
TO 2 0 m A
2 0 m A
C A L IBR AT IO N
E N TE R M ET E R
VA LU E 20 .0 00
A P P LY 0 m V
S O U R C E
C A L IB R AT IO N
C O M P LE TE
T 01 9 10 A
CALIBRATION
B - 5