Set Outputt
Key
C A LI-
B R ATE
E N TE R
[ 3 TIMES ]
E N TE R
E N TE R
E N TE R
E N TE R
E N TE R
Display
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→ YES
CALIBRATION
SENSOR TRIM
D-TO-A ADJUST
PV BIAS
→ SET OUTPUT %
SET OUTPUT %
→ LOW
HIGH
OP %: nnn.nn %
PV VAL:
value units
Hit ENTER to set OP%
ENTER NEW VALUE
value %
ID TAGNAME
READY
ADDENDUM
Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select YES .
Select SET OUTPUT %.
If digitally configured, the D-TO-A ADJUST selection
will not appear.
Select LOW and press
ENTER
HIGH selection is identical).
Display indicates present data.
.
Enter LOW value,
ENTER
. (Procedure for