Table N-1. Problem Codes of TB82 Transmitter
Code
Description
BA.CHKS Incorrect or missing 3-kΩ
Balco temperature sensor
checksum
BA.F.CAL Out of range or missing
factory calibration for 3-kΩ
Balco temperature sensor
BAD.SEE Bad SEEPROM or
pH/ORP/pION input PCB
assembly
BLNK.UP Blank microprocessor
EEPROM
GND LP
Ground loop present or
shorted sensor cable
HI.C.CKT Cable diagnostic circuit
failure - high range error
HI.CA.AD Cable diagnostic signal
above transmitter A/D range
HI.EFF
High sensor efficiency
(>110%)
Corrective Action
Cycle transmitter power.
Remove transmitter from installed location and relocate to
noise-free environment. If problem does not appear, transmitter
needs new final location or additional shielding on transmitter
and/or wiring.
Contact ABB.
Contact ABB for factory calibration procedure. Calibrate temper-
ature sensor for short-term usage until factory calibration can be
performed.
Input PCB assembly factory calibration constants can not be
loaded. Calibrate sensor and order replacement pH/ORP/pION
input PCB assembly. Existing assembly should properly function
until new assembly is received.
Cycle transmitter power.
Contact ABB.
Verify sensor wiring connections.
Verify sensor does not have any exposed wire from nicks, etc. If
it does, repair (if possible) or replace.
Remove any liquids, oils, scales or corrosion from transmitter ter-
minal block or extension cable junction box terminals.
Verify sensor responds to pH buffers. Replace sensor and/or
sensor extension cable (if present) if sensor does not respond.
Electronically test sensor (refer to the product instruction).
Replace if it does not meet requirements.
Verify sensor wiring connections.
Remove any liquids, oils, scales or corrosion from transmitter ter-
minal block or extension cable junction box terminals.
Electronically test sensor. Replace if it does not meet
requirements.
Input PCB assembly diagnostic circuit failure exists. Disable
diagnostics and order replacement input PCB assembly. Existing
input PCB assembly should properly function until new assembly
is received.
Input PCB assembly diagnostic circuit failure exists. Disable
diagnostics and order replacement input PCB assembly. Existing
input PCB assembly should properly function until new assembly
is received.
Verify proper buffer values were used for calibration and repeat
buffer calibration.
Clean sensor and repeat buffer calibration.
PROBLEM AND ERROR CODES (TB82)
TB82 TRANSMITTER
N - 17