ABB STT04 Instruction Manual page 256

Smart transmitter terminal
Table of Contents

Advertisement

Key
Display
DESCRIPTOR:
E N TE R
←PREVIOUS
DATE:
DAY: nn
E N TE R
MONTH: nn
YEAR: nn
CONFIG TYPE
600T
→ 600T EN
E N TE R
OUTPUT TYPE
→ LINEAR
E N TE R
SQU (x)
SQR (x^3)
SQR (x^5)
5th ORD. POLY
DOUBLE POLYN
DAMPING:
(0 - 16 SEC)
E N TE R
0.5 SECS
ENGINEERING UNIT
→ iH20-20c
E N TE R
O
iHg-0
ftH2O-20c
mmH20-20c
600T & 600T EN PRESSURE TRANSMITTER
Type a descriptor string using up to 16 characters.
This field can be used for notations about the device
or process.
NEXT→
To select a character, press the key that has the
desired character. Continue to press the key until the
desired character appears. Use the right arrow key to
move to the next character. Use the left arrow key to
go back to the previous character.
Enter a day and press
press
This date can represent the creation date of the con-
figuration, the date the device or element was
installed, or some other significant date.
Select 600T EN
Select LINEAR.
NOTE: Other output type selections are:
SQUARE ROOT
SQR (x^3)
SQR (x^5)
5th ORDER POLYNOMIAL
DOUBLE POLYN
Use
BACK
screen from any screen in the configuration process.
Enter a value between 0 and 16 seconds.
Select an engineering unit best suited for the
application. Other units not shown include
mmHg-0
c
PA, KPA, torr-0
mmH2O-4
Comments
. Enter a month and
ENTER
. Enter a year and press
ENTER
to return to a previous configuration
o
C, PSI, BARS, mBAR, gSqCm, Kgcm2,
o
C, ATM, MPa, iH2O-4
o
C.
CREATE/MODIFY CONFIGURATION
.
ENTER
o
C,
Q - 15

Hide quick links:

Advertisement

Table of Contents
loading