Pv Bias - ABB STT04 Instruction Manual

Smart transmitter terminal
Table of Contents

Advertisement

Key
Display
CALIBRATION
SENSOR TRIM
D-TO-A ADJUST

→ PV BIAS

SET OUTPUT %
E N TE R
PV BIAS
→ RESET
E N TE R
SET PV ZERO
SET PV VAL
PV BIAS
RESET
→ SET PV ZERO
SET PV VAL
E N TE R
PV VALUE READ:
value units
E N TE R
PRESS ENTER
TO SET PV ZERO
ID TAGNAME
READY
E N TE R
600T & 600T EN PRESSURE TRANSMITTER
Comments
Select PV BIAS.
If digitally configured, the D-TO-A ADJUST selection
will not appear.
Use the arrow keys to scroll to the desired PV BIAS
parameter. RESET removes any previously config-
ured bias values. The following procedure is used to
establish the zero offset for SET PV ZERO, the pro-
cedure for SET PV VAL is similar.
Apply the desired zero pressure value to the trans-
mitter. Scroll to SET PV ZERO using the down arrow
key and press
ENTER
.
Pressing
ENTER
calibrates the PV ZERO value.
SET PV ZERO is complete.
CALIBRATION
Q - 25