AW T4 2 0 | U N I V E R S A L 4 - W I R E , D U A L- I N P U T T R A N S M I T T E R | O I/A W T4 2 0 - E N R E V. B

6 Easy Setup

When the transmitter is started up for the first time, or when
Restore defaults is selected from the Configuration/Device
Setup/Initial Setup menu (see page 38), the 'Easy Setup'
prompt is displayed:
AWT420
03-09-2019
08:14:45
New TX
Detected
Start Easy
Setup ?
Press the
key (
) to start Easy Setup or press the
to cancel and exit to the main Operator page.
Press the
key (Edit) to change the default value/setting to
the required value/selection. Press the
the default or revised value/selection and advance to the next
parameter.
Transmitter parameters that can be configured in this way are:
Language, Instrument Tag, Diagnostics View, Signals View,
Chart View, Alarm View, Analog OP View, Calibration Log,
Alarm Log, Audit Log, Diagnostics Log, Date Format and Date
& Time. On completion of Easy Setup, the display returns to the
Easy Setup start screen:
Menu
Easy Setup
Exit
Select
Press the
key (Select) to revise/amend the settings just
made or press the
key (Exit) to cancel and exit to the main
Operator page.
All transmitter parameters can be revised/changed at any time
by selecting Enter Configuration from any Operator or View
page menu, followed by Advanced from the Access Level menu.
NOTICE
• If Easy Setup does not detect a key press within 5
minutes, the display changes automatically to the main
Operator page.
• Refer to page 25 for details of menu navigation and
parameter selection/adjustment.
7
key (
)
Ensure all electrical connections have been made correctly and
switch on the power to the transmitter. If the sensor is being
key (Next) to accept
commissioned for the first time, sensor calibration and set-up
is recommended for best results.
Refer to the following pages for calibration and setup
procedures:

Calibration and sensor setup

CAUTION
Do not attempt to setup the transmitter unless both the
sensor and transmitter are fully installed and ready for
operation.
NOTICE
• The menu structure, general operation and menu
descriptions are detailed on page 36.
• Refer to page 25 for details of menu navigation and
parameter selection/adjustment.
• 2-electrode conductivity sensors – page <?>
• 4 -electrode conductivity sensors – page 57
• pH/Redox/ORP sensors – page 59
• Turbidity/Suspended solids – page 70
• Dissolved oxygen sensors – page 81
21