Reset to Factory Configuration

Key
OUTPUT WILL BE
AFFECTED! PROCEED?
C A LI-
B R ATE
→ YES
CALIBRATION
E N TE R
→ RESET CAL
E N TE R
ID TAGNAME
This function resets all of the device parameters to the factory set-
tings. Refer to the product instruction for information on the factory
settings.
Display
NO
PROCESS CAL
EDIT CAL
WORKING
CALIBRATION
COMPLETE
READY
TBN480 CONDUCTIVITY TRANSMITTER
Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select YES.
Select RESET CAL.
CALIBRATION
J - 11