ARROW KEY ADJUSTMENT

Use this function to adjust the four to 20 milliampere output of the field
device using the up and down arrow keys.
Key
Display
OUTPUT WILL BE
AFFECTED! PROCEED?
C A LI-
NO
B R ATE
→ YES
CALIBRATION
→ D-TO-A ADJUST
E N TE R
BENCH CALIB.
D/A CAL USING
→ UPDOWN ARROW KEYS
E N TE R
METER VALUE ENTRY
ADJUST TO 4 mA
E N TE R
THEN HIT ENTER
ADJUST TO 20 mA
E N TE R
THEN HIT ENTER
ID TAGNAME
E N TE R
READY
PTH PRESSURE TRANSMITTER
Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select YES.
Select D-TO-A ADJUST.
Select UPDOWN ARROW KEYS.
Use the arrow keys to adjust the 4 mA signal.
NOTE: When increasing or decreasing the mA sig-
nal, the increments of change increase with succes-
sive depressions until the maximum level of change
is reached. By changing direction you will return to
the smallest increment of change. This adjustment
technique speeds up the adjustment process without
affecting fine adjustment.
Use the arrow keys to adjust the 20 mA signal.
CALIBRATION
D - 9