TBN580 TRANSMITTER
Process Calibration
Key
OUTPUT WILL BE
AFFECTED! PROCEED?
C A LI-
B R ATE
→ YES
CALIBRATION
→ PROCESS CAL
E N TE R
BENCH CAL pH
BENCH CAL TEMP
D-TO-A ADJUST
INSTALLING A
NEW SENSOR?
E N TE R
→ NO
TEMP COMP TYPE
IN USE:
E N TE R
ENTER FIXED
PROCESS TEMP
E N TE R
CHANGE TEMP
OF PROCESS?
E N TE R
CALIBRATION
K - 6
This procedure applies to TBN transmitters configured for either ANA-
LOG or DIGITAL. Any differences between the two are noted in the
Comments column.
Display
NO
YES
HIT ENTER
nnn.n°UNITS
nnn°UNITS
Comments
Select YES.
Select PROCESS CAL.
If configured digitally, the D-TO-A ADJUST selection
would not appear.
Select NO.
If YES, a new sensor offset will be calculated and the
cumulative offset change will be reset to 0.00 mV.
Displayed will be the TEMP COMP TYPE selected in
configuration.
If TEMP COMP TYPE is MANUAL or NONE (as
selected in configuration), this screen will appear,
otherwise, the next screen, CHANGE TEMP OF
PROCESS? will appear. Default is 25°C.
The value displayed is the temperature of the pro-
cess (interior of sensor).
NOTE: If the new temperature value is entered, this
change will affect the calibration of temperature.