ABB STT04 Instruction Manual page 213

Smart transmitter terminal
Table of Contents

Advertisement

Key
Display
ENTER PROCESS
TEMPERATURE
E N TE R
nn.nn
LOW CALIB
E N TE R
HIGH CALIB
APPLY
E N TE R
PRESS ENTER
WHEN STABLE
APPLY
E N TE R
PRESS ENTER
WHEN STABLE
CALCULATED
EFFICIENCY
E N TE R
nn.n %
PRESS ENTER
CALCULATED
SENSOR OFFSET
E N TE R
nnn.n mV
PRESS ENTER
RELEASE OUTPUT HOLD
NO
E N TE R
→ YES
ID TAGNAME
E N TE R
READY
Enter a process temperature value.
This screen appears only for 2 POINT CAL when
performing a pH CAL. 25
C
O
Enter the low value of the buffer or standard and
press
nn UNITS
The low default value is 4.0 pH. The high default
nn UNITS
value is 7.0 pH.
Apply the low value of the buffer or standard.
n.nn UNITS
Apply the high value of the buffer or standard.
n.nn
UNITS
The terminal displays the calculated efficiency in per-
centage.
The terminal displays the calculated sensor offset in
millivolts.
Select YES to release output hold. Screen appears
at the end of two-point calibration when device is in
the analog mode.
If any warnings or messages exist pertaining to this
device, the terminal will list them.
TB82 TRANSMITTER
Comments
°
C is the default value.
. Enter the high value.
ENTER
CALIBRATION
N - 7