Calibration; Process Calibration; Bench Calibration (Orp/Pion - ABB STT04 Instruction Manual

Smart transmitter terminal
Table of Contents

Advertisement

TBN581 (ORP/pION) TRANSMITTER
Key
ENTER
SECONDARY UNITS
E N TE R
_nnnnnn
SECONDARY L.R.
E N TE R
SECONDARY U.R.
STORE THIS
CONFIGURATION
E N TE R
→ YES
ID TAGNAME
E N TE R

CALIBRATION

CALIBRATION
L - 4
Display
_nn.nn UNITS
nn.nn UNITS
NO
READY
This section details the calibration functions of the STT04 terminal
while interfacing a TBN581 ORP/pION transmitter. There are four
types of calibration functions:

Process calibration.

Bench calibration (ORP/pION).

Bench calibration (mV).
D-to-A adjust.
NOTES:
1.
When calibrating, a field device must be connected to the
STT04 terminal.
2.
To change calibration or configuration parameters of a field
device that is connected to an IMFBS01 module, the transmitter
must be taken off-line. This is done at the INFI 90 OPEN console.
Refer to Figure
L-2
Comments
Enter up to 6 characters. Use familiar units to
describe the output.
Input the lower range value and press
the upper range value.
Select YES to store the configuration.
The Configuration ID tag name just configured will be
in the upper left corner if the configuration is saved. It
becomes the working configuration.
NOTE: Connect the terminal to a device and send
the configuration (using
SEND CONFIG
before calibrating.
for an overview of the calibration functions.
. Input
ENTER
) to a device