Single-Point Calibration - ABB 4600 Series Operating Instructions Manual

Oxygen transmitters,for use with zgp2 probes
Table of Contents

Advertisement

...7 PROGRAMMING
Note. Before commencing a gas calibration procedure the analyzer and probe must be switched on and allowed to run
for at least one hour to allow the system to stabilize thermally.

7.1 Single-point Calibration

The calibration sequence involves standardizing the analyzer and the oxygen probe, using air as the test gas. Until a calibration
sequence has been completed successfully, the existing slope remains unaffected.
-----
OXYGEN CAL. SEQ.
0. 0
Cell Constant mV
100. 0
Span % of Theory
0000 . 0
Cal. User Code
-----
One Point Cal.
Two Point Cal.
Preset Cal.
20. 9 5
Connect to Air
20. 9 5
Calibrating Air
-----
Calibration Pass
Failed Constant
Failed Unstable
SECURITY CODE
14
Oxygen Calibration Sequence
Cell Zero mV
The upper display shows the millivolt offset of the oxygen probe from the last
successful calibration.
Span % of Theory
A value between 90 and 110% should be displayed. The display indicates the
oxygen probe output slope using parameters derived at the last successful
two-point calibration or the preset values.
Calibration User Code
[00000 to 19999]
If an incorrect value is entered, access to the calibration sequence is inhibited
and the display returns to the beginning of the
One Point Calibration
Select one-point calibration sequence.
Connect to Air
Connect the air supply to the probe (refer to the probe's Operating Instructions,
IM/ZGP2 ). The upper display indicates the measured % oxygen.
Calibrating Air
The upper display indicates the measured % oxygen. When a stable reading is
detected the display advances automatically to the next frame. To abort
calibration, press either the
Calibration Pass/Fail
On completion a calibration status message is displayed:
Calibration Pass: Calibration sequence successful
Failed Constant:
Failed Unstable:
Note. If the sensor calibration is unsuccessful then the
parameters are unaffected. The instrument continues to
Span % of Theory
operate using parameters stored during the last successful calibration.
Press
to return to the beginning of the
Press
to advance to the
Oxygen Calibration Sequence
or
switch.
Cell offset >±10mV
(upper display shows cell mV output)
Cell output unstable (drifting).
Oxygen Calibration Sequence.