Reset Calibration - ABB STT04 Instruction Manual

Smart transmitter terminal
Table of Contents

Advertisement

TB82 TRANSMITTER

Reset Calibration

Key
OUTPUT WILL BE
AFFECTED! PROCEED?
C A LI-
B R ATE
→ YES
E N TE R
→ RESET CAL
OUTPUT WILL BE
AFFECTED! PROCEED?
E N TE R
→ YES
ID TAGNAME
E N TE R
CALIBRATION
N - 10
This procedure describes how to reset a calibration back to the
default settings.
Display
NO
nnnn CAL
TEMP. CAL
EDIT CAL
NO
READY
Comments
This operation will cause a change in output not cor-
responding to the input. Selecting YES will hold the
output during calibration.
Select YES.
Select RESET CAL.
This function sets all process sensor and tempera-
ture sensor calibration data to the default settings.
The following are the factor settings.
Sensor slope: 100 %
Sensor offset: 0 mV
Temperature slope: 1
Temperature offset: 0°C
Select YES to reset the calibration.