Edit Calibration - ABB STT04 Instruction Manual

Smart transmitter terminal
Table of Contents

Advertisement

Edit Calibration

Key
OUTPUT WILL BE
AFFECTED! PROCEED?
C A LI-
B R ATE
→ YES
E N TE R
→ EDIT CAL
PROCESS
E N TE R
SLOPE :
OFFSET:
TEMPERATURE
E N TE R
SLOPE :
OFFSET:
ID TAGNAME
E N TE R
This procedure describes how to edit a calibration.
Display
NO
nnnn CAL
TEMP. CAL
RESET CAL
nnn.nnnn
nnnn.nnnn
nn.nnnn
nnn.nnnn
READY
TB82 TRANSMITTER
Comments
This operation will cause a change in output not cor-
responding to the input. Selecting YES will hold the
output during calibration.
Select YES.
Select EDIT CAL.
Enter the sensor slope value and press
Valid values are 40% to 150%.
Enter a sensor offset value. Valid values are -1,000
to +1,000 mV.
Enter a temperature slope value and press
Valid range is 0.2 to 1.5.
Enter a temperature offset value. Valid range is -40°
to 40°C.
If any warnings or messages exist pertaining to this
device, the terminal will list them.
.
ENTER
.
ENTER
CALIBRATION
N - 9