Replacement Procedure

A1 Freq / Power Detector Module Replacement

A1 Freq / Power Detector Module Replacement

Tools Required

TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 1.02 N-m / 9.0 lb-in) (for reconnecting SMA
connector)

Removal Procedure

Refer to
Figure 6-17
Step 1. Remove the outer cover as described in
Step 2. Remove the front panel as described in
Step 3. Disconnect the RF cable(item 1).
Step 4. Disconnect the semi-rigid cables(item 2).
Step 5. Disconnect the semi-rigid cable(item 3).
Step 6. Disconnect the semi-rigid cable(item 4).
Step 7. Remove the TORX T15 screw fastening the freq / power detector module bracket (item 5).
Step 8. Slide out the freq / power detector module (item 6).
168
for this procedure.
"Outer Cover Removal" on page 136
"Front Panel Removal" on page
138.