Replacing The Inner Source Assembly Filament; Replacing The Filament - Waters Xevo TQ-GC Overview And Maintenance Manual

Table of Contents

Advertisement

Figure 4–23: Correctly fitted EI and CI filament housing assemblies
EI filament housing assembly fitted with two yttrium oxide-coated rhenium EI
filaments
EI Filament 1
EI Filament 2
CI filament housing assembly fitted with one uncoated rhenium CI filament
5.
Fit the inner source into the instrument (see

4.8 Replacing the inner source assembly filament

Replace the inner source assembly filament when either of the following conditions apply:
• If there is no ion beam, and the readbacks for emission current and filament current are 0, a
damaged filament is usually the cause.
• Poor sensitivity indicates that the filament may be old and inefficient.
See also
Troubleshooting the Xevo

4.8.1 Replacing the filament

Required materials:
• Chemical-resistant, powder-free gloves
• 2.5-mm hex wrench (supplied with the instrument)
TQ-GC.
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
Fitting the inner source to the
Page 73
instrument).