5.23 Replacing the ion block source heater
Replace the ion block source heater if it fails to heat the ion block when the instrument is
pumped-down (evacuated).
Required materials
• Chemical-resistant, powder-free gloves
• Needle-nose pliers
• Combined 2.5-mm hex wrench and cone extraction tool
• New ion block source heater assembly
To replace the ion block source heater:
1. Remove the ion block assembly from the instrument (see
assembly from the source assembly (Page
2. Ensure that the isolation valve is closed.
Figure 5–94: Ion block assembly
Source isolation valve handle in closed position
Sampling cone assembly retaining blocks
Cone gas nozzle handle
3. Use the combined 2.5-mm hex wrench and cone extraction tool to loosen the two captive
screws securing the ion block cover plate.
73)).
March 17, 2021, 715005046 Ver. 03
Page 140
Removing the ion block