To fit the new IonSABRE II probe heater:
Caution:
electrical connections, capillary sleeve, or capillary when fitting
the heater over the capillary sleeve.
1.
Use the probe adjuster knob to adjust the capillary so that it protrudes
approximately 0.5 mm from the end of the probe.
2.
Carefully slide the probe heater over the capillary sleeve on the probe
assembly.
Caution:
connections, do not twist the heater when fitting it to the probe
assembly.
3.
Fit the probe heater to the probe assembly, ensuring that the heater is
fully seated on the probe assembly.
4.
Fit the probe to the instrument (see
5.
In the Instrument Console, click API
Replacing the ion block source heater
Replace the ion block source heater if it fails to heat when the instrument is
pumped down (evacuated).
Required materials
•
Chemical-resistant, powder-free gloves
Take great care not to damage the probe heater's
Probe heater connections
To avoid damaging the probe heater's electrical
0.5 mm
Capillary
page
3-6).
to start the desolvation gas.
Replacing the ion block source heater
Capillary sleeve
4-87