Replacing The Ion Block Source Heater - Waters Xevo G2 QTof Operator's, Overview And Maintenance Manual

Hide thumbs Also See for Xevo G2 QTof:
Table of Contents

Advertisement

Replacing the ion block source heater

Replace the ion block source heater if it fails to heat the ion block when the
instrument is pumped-down (evacuated).
Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
Combined 2.5-mm Allen wrench and cone extraction tool
New ion block source heater
To replace the ion block source heater:
1.
Remove the ion block assembly from the instrument (see
2.
Ensure that the isolation valve is closed.
5-94
Maintenance Procedures
The ion block assembly can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
page
5-38).
Isolation valve handle
in closed position