Replacing fluid lines

If the inlet, outlet, or waste lines become damaged, replace them to ensure
optimal performance. Each fluid line is supplied with a pre-fitted PEEK™
connector assembly at the point where it connects to the TRIZAIC source
module (see
capillary tubing and the PEEK assembly as a single unit. No parts of the fluid
line are reusable once detached.
Fluid line connections:
To identify the correct tubing assembly for each fluid line, use the following
table.
TRIZAIC tubing assemblies:
Part Number
430002561
430002351
430002455
You can use these part numbers to identify each line when it is
Tip:
disconnected from the source. Part numbers are embossed on the shrink-wrap
below the ferrule assembly.
3-4
Maintenance Procedures
figure "Fluid line connections:" on page
O-ring
Capillary tubing
Description
Infusion line (Orange)
Inlet line (Orange)
Waste line (White)
Connector plate
Compression
fitting
PEEK fluid
connector
3-4). You must replace the