Operation
Instrument Controls Menu
3-18
Model 5028i Instruction Manual
In the Main Menu, choose Instrument Controls > Filter Tape
l
Control.
The Manual screen allows the user to move or stop the filter tape and

Manual

open or close the measurement bench.
Note Channel B settings please refer to Channel A as following
description.
In the Main Menu, choose Instrument Controls > Filter Tape
l
Control > Channel A or Channel B>Manual.

Mass Limit

The Mass Limit screen allows the user to set the mass limit. The default
is 1,500 µg (or 1.5 mg) and is specified as part of any U.S. EPA and
TUV approvals. However, the limit can be increased as much as 5,000
µg for non-regulatory monitoring.
Note Channel B settings please refer to Channel A as following
description.
In the Main Menu, choose Instrument Controls > Filter Tape
l
Control > Channel A orChannel B> Mass Limit.
FILTER TAPE CONTROL:
>CHANNEL
A
CHANNEL
B
RANGE
AVG
CHANNEL:
>MANUAL
MASS LIMIT

NEXT TIME

PERIOD
COUNTER
RANGE
AVG
s
MANUAL:
>BENCH(S)
TAPE
RANGE
AVG
s
DIAGS ALARM
µ
1,500
g
05MAY08 00:00
8 HRS
0
DIAGS ALARM
CLOSED
STOP
TOGGLE VALUE
DIAGS ALARM