Thermo Scientific 5028i Instruction Manual page 277

Continuous particulate monitor
Table of Contents

Advertisement

Thermo Fisher Scientific
set thermal mass coef a/b value
This command sets the thermal mass coefficient to user-defined values
to value, where value is a floating-point representation of the coefficient.
The following example sets the thermal mass coefficient to 20.5.
Send:
set thermal mass coef a 20.5
Receive:
set thermal mass coef a 20.5 ok
Send:
set thermal mass coef b 20.5
Receive:
set thermal mass coef b 20.5 ok
vac mass coef a/b
This command reports the current vacuum mass coefficient. The
following example reports that the vacuum mass coefficient is 2.0.
Send:
vac mass coef a
Receive:
vac mass coef a 2.0
Send:
vac mass coef b
Receive:
vac mass coef b 2.0
set vac mass coef a/b value
This command sets the vacuum mass coefficient to user-defined values
to value, where value is a floating-point representation of the coefficient.
The following example sets the vacuum mass coefficient to 3.1.
Send:
set vac mass coef a 3.1
Receive:
set vac mass coef a 3.1 ok
Send:
set vac mass coef b 3.1
Receive:
set vac mass coef b 3.1 ok
vac pres span a/b
This command reports the vacuum span for the detector at the time of
calibration. The following example reports that the vacuum pressure
span is 1.0.
Send:
vac pres span a
Receive:
vac pres span a 1.000
Send:
vac pres span b
Receive:
vac pres span b 1.000
set vac pres span a/b value
This command sets the vacuum span to the value specified at the time
of calibration. The following example successfully sets the new value for
vacuum pressure span to 1.002.
Send:
set vac pres span a 1.002
Receive:
set vac pres span a 1.002 ok
Send:
set vac pres span b 1.002
Receive:
set vac pres span b 1.002 ok
C-Link Protocol Commands
Calibration
Model 5028i Instruction Manual
B-37