The RH/Temperature Calibration submenu allows the user to view and
set the relative humidity and temperature sensor calibration. The RH/
temperature calibration submenu is visible only when the instrument is
in service mode. For more information on the service mode, see
Mode" earlier in this chapter.
Note This adjustment should only be performed by an instrument
service technician.
s
In the Main Menu, choose Service > RH/Temp Calibration.
l
RH/TEMP CALIBRATION:
>AMBIENT TEMP
FLOW TEMP
FLOW TEMP B
AMBIENT RH
RANGE
The Ambient Temperature screen allows the user to view and calibrate
ambient temperature. Adjust the zero offset until the ambient
temperature agrees with your treaceable standard.
In the Main Menu, choose Service > RH/Temp Calibration >
l
Ambient Temp.
CALIBRATE AMBIENT TEMP:
TEMPERATURE:
RANGE
The Flow/Flow B Temperature screen allows the user to view and
calibrate flow/flow B temperature. Be sure to have the cover removed
and equilibrated to room temperature.
Note Flow B Temperature settings please refer to Flow Temperature as
following description.
In the Main Menu, choose Service > RH/Temp Calibration > Flow
l
Temp/ Flow Temp B.
CALIBRATE FLOW TEMP:
TEMPERATURE:
RANGE
AVG
DIAGS ALARM
OFFSET:
INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
s
OFFSET:
INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
Operation
Service Menu
"Service
o
4.2
C
o
0.8
C
o
0.8
C
0.0
%
o
29.9
C
o
4.2
C
o
28.0
C
o
0.8
C
Model 5028i Instruction Manual
3-71

Advertisement

Table of Contents
loading

Table of Contents