C-Link Protocol Commands
Datalogging
B-24
Model 5028i Instruction Manual
The following example copies the current list of lrecs into the scratch
pad.
Send:
set copy lrec to sp
Receive:
set copy lrec to sp ok
data treatment lrec
data treatment srec
These commands report the current selection of data treatment for
concentrations in the lrecs or srecs. The following example reports the
data treatment for concentrations in lrec as minimum.
Send:
data treatment lrec
Receive:
data treatment lrec min
set data treatment lrec string
set data treatment srec string
string = | cur | avg | min | max |
These commands set the data treatment to current, average, minimum,
or maximum for the concentration values recorded in the lrecs or srecs.
The following example sets the data treatment for concentrations in lrec
as minimum.
Send:
set data treatment lrec min
Receive:
set data treatment lrec min ok
erec
erxy
x = | 0 | 1 |
: Reply termination format (see "set format format"
command)
y = | 0 | 1 | 2 | : Output format (see "set erec format format" command)
These commands return a snapshot of the main operating conditions
(measurements and status) at the time the command is issued. The
following example shows a typical response.
The format is defined within the command (in the case of erxy) by the
current settings of the "format" and "erec format" commands (in the
case of erec). For details on erec formatting, see the
Definition" section at the end of this appendix. For details on how to
decode the flag fields within these records, see the "flags" command.
Send:
erec
Receive:
erec
07:53 05-04-09 flags 110000 PM 0.000 1 24Hr Avg PM
0.000 1 Coef A 1.000 Bkg A 0.000 Range A 10000.000
Flow A 0.000 6 5 3 1 1 1 0 Temp Threshold A 30.000
Temp Threshold A 35.000