Operation
Instrument Controls Menu
Screen Contrast
3-50
Model 5028i Instruction Manual
User Value
The User Value screen allows the user to set the output value for
the corresponding input voltage for the selected table point in the
conversion table, from -9999999 to 99999999. The default table is a
two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U, where U is the previously entered unit of measure.
In the Main Menu, choose Instrument Controls > I/O
l
Configuration > Analog Input Config > select Channel > select
Point > User Value.
The Screen Contrast screen is used to change the contrast of the display.
Intensities between 0 and 100% in increments of 5 are available.
Changing the screen contrast may be necessary if the instrument is
operated at extreme temperatures.
Note The optimal contrast will change with changes in temperature.
Note The optimal contrast will change from one LCD to another. If the
LCD is replaced, the contrast may need to be reset.
If the display contrast is not optimal, but the content on the screen is
visible, select Instrument Controls > Screen Contrast and adjust the
screen contrast. If the content on the screen is not visible, use the "set
contrast 10" C-Link command to set the screen contrast to mid-range,
then optimize the contrast. See "contrast levels" in Appendix B,
Protocol
In the Main Menu, choose Instrument Controls > Screen Contrast.
l
TABLE POINT 01 USER VAL:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
AVG
Commands" for more information on this command.
SCREEN CONTRAST:
CURRENTLY:
SET TO:
RANGE
AVG
0.00
00000.00
MOVE CURSOR
SAVE
DIAGS ALARM
s
50 %
55 % ?
CHANGE VALUE
SAVE VALUE
DIAGS ALARM
Thermo Fisher Scientific
s
"C-Link
s