Thermo Scientific 5028i Instruction Manual page 98

Continuous particulate monitor
Table of Contents

Advertisement

Operation
Instrument Controls Menu
Analog Output Configuration
(Select Action)
Select Range
3-44
Model 5028i Instruction Manual
ALLOW OVER/UNDER RANGE:
CURRENTLY:
SET TO:
RANGE
The Analog Output Configuration menu displays a list of the analog
output configuration choices, from which the user selects the parameter
to adjust for the selected output channel. Configuration choices include
selecting range, setting minimum/maximum values, and choosing the
signal to output.
In the Main Menu, choose Instrument Controls > I/O
l
Configuration > Analog Output Config > All Voltage Channels, All
Current Channels, Voltage Channel 1–6 or Current Channel 1–6.
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
RANGE
The Select Range screen is used to select the hardware range for the
selected analog output channel. Possible ranges for the voltage outputs
are: 0-100 mV, 0-1, 0-5, and 0-10 V. Possible ranges for the current
outputs are: 0-20 mA and 4-20 mA.
In the Main Menu, choose Instrument Controls > I/O
l
Configuration > Analog Output Config > select Channel > Select
Range.
SELECT OUTPUT RANGE:
SELECTED OUTPUT:
CURRENTLY:
RANGE
TOGGLE VALUE
AVG
DIAGS ALARM
AVG
DIAGS ALARM
SET TO:
0-100mV ?
CHANGE VALUE
AVG
DIAGS ALARM
ON
OFF
?
V ALL
0-10V
SAVE
Thermo Fisher Scientific