Instrument Alarms
Filter Tape Counter
Thermo Fisher Scientific
Items displayed are determined by the options installed. To see the
actual reading of an item and its minimum and maximum limits, move
the cursor to the item and press
the minimum or maximum limit, an alarm is activated and the alarm
(bell) icon appears in the status bar on the Run screen and in the Main
Menu.
If any alarms are active, the alarm (bell) icon is displayed on the right
side of the status bar.
In the Main Menu, choose Alarms.
l
ALARMS:
ALARMS DETECTED
>INSTRUMENT ALARMS
DETECTOR ALARMS
RH/TEMPERATURE ALARMS
PRESSURE/VACUUM ALARMS 0
FLOW ALARMS
CONC ALARMS
RANGE
The Instrument Alarms submenu allows the user to view a set of alarms
for different boards. The filter tape change, heater power, and bench
items (motherboard, interface board and I/O expansion board) are read
only. The motherboard status, interface board status, and I/O expansion
board status (if installed) indicate that the power supplies are working
and connections are successful. There are no setting screens for these
alarms.
In the Main Menu, choose Alarms > Instrument Alarms.
l
INSTRUMENT ALARMS:
>FILTER TAPE COUNTER
FILTER TAPE CHANGE
BENCH
HEATER POWER
MOTHERBOARD STATUS
INTERFACE STATUS
I/O EXP STATUS
RANGE
FILTER TAPE COUNTER B
FILTER TAPE CHANGE B
BENCH B
HEATER POWER B
The Filter Tape Counter screen displays the current filter tape counter
value and sets the maximum alarm limit. If the filter tape counter
reading goes beyond the maximum limit, an alarm is activated and the
alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu.
. If the readings go beyond either
AVG
DIAGS ALARM
AVG
DIAGS ALARM
Model 5028i Instruction Manual
Operation
Alarms Menu
0
0
0
0
0
0
OK
OK
OK
OK
OK
OK
OK
3-59