Thermo Scientific 5028i Instruction Manual page 91

Continuous particulate monitor
Table of Contents

Advertisement

TCP/IP Settings
Thermo Fisher Scientific
The TCP/IP Settings menu is used for defining parameters that are
required for Ethernet communications.
Note The instrument power must be cycled after any of these
parameters have been changed for the change to take effect.
In the Main Menu, choose Instrument Controls > Communication
l
Settings > TCP/IP Settings.
Use DHCP
The Use DHCP screen is used to specify whether to use Dynamic Host
Configuration Protocol (DHCP) or not. When DHCP is enabled, the
network dynamically provides an IP address for the instrument.
In the Main Menu, choose Instrument Controls > Communication
l
Settings > TCP/IP Settings > Use DHCP.
IP Address
The IP Address screen is used to edit the IP address. The IP address can
only be changed when DHCP is off. If DHCP is on, the instrument
will respond with "NOT SETTABLE IF DHCP IS ON". For more
information on DHCP, see
In the Main Menu, choose Instrument Controls > Communication
l
Settings > TCP/IP Settings > IP Addr.
TCP/IP SETTINGS:
>USE DHCP
IP ADDR
10.68.179.67
NETMASK
255.255.254.0
GATEWAY
HOST NAME
NTP SVR
RANGE
AVG
DIAGS ALARM
DHCP:
CURRENTLY:
SET TO:
TOGGLE VALUE
CYCLE POWER TO CHANGE DHCP
RANGE
AVG
DIAGS ALARM
"Use
DHCP" above.
IP ADDRESS:
CURRENT: 10.68.179.67
SET TO: 10.68.179.67
MOVE CURSOR
CHANGE VALUE
SAVE VALUE
RANGE
AVG
DIAGS ALARM
Instrument Controls Menu
OFF
10.68.178.1
iSeries
0.0.0.0
OFF
ON ?
Model 5028i Instruction Manual
Operation
s
3-37